Product Overview
Thermal conductivity reading system with a measurement area of at least 600 × 600 mm for boards, loose-fill and compressible specimens. The specimen sits between two temperature-controlled plates; the steady heat flow is measured and thermal conductivity is calculated from the plate temperature difference, specimen thickness and heat flow sensor data. Covers 0.002–2 W/(m·K) at −20 °C to 90 °C.
Core Features
- Measurement area of at least 600 × 600 mm
- Thermal conductivity range: 0.002–2 W/(m·K)
- Measurement temperature: −20 °C to 90 °C
- Repeatability ±0.25 %, temperature stability ±0.05 °C
- Motorized upper plate with parallel, repeatable specimen contact
- Controlled compression at 5 kPa and 10 kPa contact pressure for TS EN 12667
- Target thickness regulation: upper plate moves automatically to the set specimen thickness
- Low-resistance specimens below 0.3 m²K/W with external thermocouples and compensation mats
- PC + touchscreen regulation, automatic Excel / Word / PDF reports, remote service
Standards
- TS EN 12667 — Thermal capability of building substances and products — determination of thermal resistance by means of guarded hot plate and heat flow meter methods — products of high and medium thermal resistance
- TS EN 12664 — Thermal performance of building materials and products — determination of thermal resistance by means of guarded hot plate and heat flow meter methods — products of medium and low thermal resistance
- TS 415 EN 12939 — Thermal capability of building substances and products — thick products of high and medium thermal resistance
- ASTM C518 — Steady-state thermal transmission properties by means of the heat flow meter apparatus
- ISO 8301 — Thermal insulation — determination of steady-state thermal resistance and related properties — heat flow meter instrument
Specifications
- Thermal Conductivity Range0.002–2 W/(m·K)
- Minimum Reading Area600 × 600 mm
- Reading Temperature−20 °C … 90 °C
- Plate Setpoint Range−10 °C … 60 °C
- Accuracy (vs. reference substance)±1 % – ±2 %
- Accuracy with Calibration SpecificationBetter than ±1 %
- Repeatability±0.25 %
- Reproducibility±0.5 %
- Temperature Stability±0.05 °C or better
- Setpoint Accuracy±0.03 °C or better
Heating Time (10 °C rise): < 10 min (typical)
- Heating CapacityMin. 2,000 W
- Cooling CapacityMin. 840 W (at 20 °C)
- Contact Pressure5 kPa; 10 kPa (TS EN 12667)
- Calibration ReferencesEPS specification 600 × 600 × 25 mm; ERM-FC440C 600 × 600 × 28.7 mm
- SoftwareWindows 10/11 (64-bit), up to four devices on one PC, QS/ERP integration
Specifications
| Parameter | Specification |
|---|---|
| Model | RTPLINE-AL-λ 600 |
| Brand | RTPLINE |
| Code | 9024800000 |
Specifications are configurable to your pipe sizes and test standard. Ask us for the full datasheet and a configuration proposal.