Product Overview
Dedicated lab solution for testing the surface-temperature resistance and capability of rendered mortar specimens under controlled infrared heating. Four independent examine lines each have an insulated closed chamber and a top-mounted infrared heater about 400 mm above the specimen; the surface is held at 60 ± 2 °C for 8 hours ± 15 min while all lines are monitored and recorded from a single 10" touchscreen.
Key Features
- 4 independent examine lines, each with its own insulated closed chamber
- Specimen size 400 × 400 × 40 mm, loaded horizontally on a slide-out rail tray
- Top-mounted infrared heater, about 400 mm above the specimen surface
- Surface temperature 60 ± 2 °C held for 8 hours ± 15 min
- One PT100 or K-type thermocouple per specimen, accuracy ±0.5 °C or better
- Single 10" touchscreen for control, monitoring and reporting; USB export of records
- Resumes from where it stopped after a power cut
- Supply: 380 V AC three-phase or 220–230 V AC single-phase (optional)
- PC and reporting software included; 2-year warranty and 10-year service and spare parts support
Standards
- TS EN 998-1 — Specification for mortar for masonry — rendering and plastering mortar
- TS EN 1015-21 — Methods of examine for mortar for masonry — determination of the compatibility of one-coat rendering mortars with substrates
Specifications
- Standard ComplianceTS EN 998-1; TS EN 1015-21
- Examine Lines4 independent lines
- Specimen Size400 × 400 × 40 mm
- Surface Temperature60 ± 2 °C
- Temperature Hold Time8 hours ± 15 min
- HeatingInfrared, top-mounted, about 400 mm to specimen surface
- Temperature SensorsPT100 or K-type thermocouple, one per specimen
- Measurement Accuracy±0.5 °C or better
- Data Recording IntervalMax. 5 min
- Data ExportUSB (at least 1 month of records)
- After Power FailureContinues from where the examine stopped
- Regulation10" touchscreen, microprocessor-controlled
- Supply380 V AC 3-phase + N + PE, 50–60 Hz (220–230 V AC single-phase optional)
- WarrantyAt least 2 years
FAQ
- QHow many specimens can be tested at once and what size are they?
- AFour specimens at the same time, one per independent line, each 400 × 400 × 40 mm. Specimens are loaded horizontally on a slide-out rail tray.
- QWhat temperature and test duration does it offer?
- AA constant surface temperature of 60 ± 2 °C held for 8 hours ± 15 minutes, measured by one PT100 or K-type sensor per specimen with ±0.5 °C accuracy or better.
- QHow is the data recorded and what happens in a power cut?
- ARecords are taken at intervals of at most 5 minutes and can be exported by USB (at least one month of records). After a power cut the examine continues from where it stopped. A PC and reporting software are included.
- QWhat are the power supply options, service and warranty?
- A380 V AC three-phase, or 220–230 V AC single-phase as an option. Installation, commissioning and user training are provided, with at least 2 years of warranty and 10 years of service and spare parts support.
Specifications
| Parameter | Specification |
|---|---|
| Model | RTPLINE-4-Line |
| Brand | RTPLINE |
| Code | 9024800000 |
Specifications are configurable to your pipe sizes and test standard. Ask us for the full datasheet and a configuration proposal.
Source: RTPLINE product engineering data · Updated November 2026